BMT Incucell V 222 ECO, Forced Air Circulation Incubator, 7.8 cu. ft. (222L) 115v

BMT Incucell V 222 ECO, Forced Air Circulation Incubator, 7.8 cu. ft. (222L) 115v

BMT Incucell V 111 ECO, Forced Air Circulation Incubator, 3.9 cu. ft. (111L) 115v

BMT Incucell V 111 ECO, Forced Air Circulation Incubator, 3.9 cu. ft. (111L) 115v

BMT Incucell V 222 ECO, Forced Air Circulation Incubator, 7.8 cu. ft. (222L) 115v

Availability: In stock
SKU
MC 000230
Description

Forced Air Circulation Incubator

$6,040.00
BMT Incucell V 222 ECO, Forced Air Circulation Incubator, 7.8 cu. ft. (222L) 115v is available to buy in increments of 1

INCUCELL Natural, precise

Suitable for safe treatment of microbiological cultures. The INCUCELL line produces no noise and provides a very soft air flow within the chamber. These devices can be used especially in biological and microbiological laboratories, quality tests in pharmacy, cosmetic industry and testing in veterinary medicine and food processing industry.

The patented air flow system ensures a rapid and precise reaching of the required temperature throughout the chamber and consequently the ideal process of the samples heating. The INCUCELL V series offers high operation comfort, precise control and short temperature recovery times in the chamber after opening the door and therefore it is suitable for demanding and precise tests and cultivation processes. The INCUCELL V devices meet requirements of technical-legislative regulations in CZ and EU.

The cost-effective incubator series INCUCELL with the chamber volume of 22, 55, 111, 222, 404, and 707 liters, working temperature from +5 °C above ambient temperature up to 100 °C and precise temperature control is suitable for high-quality and safe processing of microbiological cultures in laboratories and industry. The series INCUCELL is characterized by noiseless run and soft air flow. It offers high operating comfort, precise control and short times of temperature recovery in the chamber after the door opening. The devices INCUCELL meet requirements of technical-legislative regulations both in CZ and EU.

Sturdy design and perfect protection against corrosion are ensured by a self-supporting outer jacket made of galvanized steel sheet with hard and resistant baking varnish on the outer side. The inner double-jacket chamber and the door panel are made of high-quality stainless steel (DIN 1.4301, AISI 304). For the purposes of energy saving and minimizing the heat radiated to the ambient air, the chamber is perfectly insulated with an ecological insulating material. The inner walls of the chamber can be easily removed without using any tool, e.g. for the purposes of cleaning. On top of that the chamber has a continuously adjustable air flap and inner glass door. The door is secured at four points for perfect tightening.

Features:

  • Natural, precise – laboratory incubator
  • 55-liter chamber volume
  • Working temperature of from +10°C over ambient temperature up to 100°C
  • Patented system of air flow in the chamber ensures quicker and more accurate distribution of the preset temperature in the chamber, which enables ideal heating of the samples
  • High operation comfort, precise temperature regulation and short times of temperature recovery in the chamber after the door opening; the standard equipment includes inner glass door
  • Model with a fan ensures quicker and more accurate distribution of temperature with small deviations; suitable for demanding and accurate tests and samples cultivation
  • Standard model with a Fuzzy logic microprocessor control system; control elements situated on a foil keyboard, process information indicated on a LED display
  • Wide range of optional equipment and accessories
  • Output for a printer or PC connection
  • Validation possibility (IQ , OQ) 

 

High-tech systems for incubation – both heating and cooling, humidity and light control – intended for use in research, development, production, business activities, or quality control. The sectional principle that applies to the design of laboratory ovens allows a wide range of optional equipment and accessories.

The basic model is available – depending on the oven type – in volume size range from 22 to 1212 liters, with either natural or forced air circulation, natural heating and cooling, controlled humidity and controlled CO2 atmosphere. The incubators are produced in two levels of operation comfort – Standard and Comfort line.

The parameters inside the unit chamber are monitored and documented with the help of a PC, using a special SW WarmComm (optional equipment).

The laboratory incubators line meets the requirements of reproducibility and accuracy of temperature distribution and other parameters in laboratory incubators, derived form such quality standards like ISO 9000 or ISO/17025.

Our laboratory incubators comply with the requirements of operation in accredited or certified laboratories

Incu

 

 

Vacucell

 

eco

 

 

Technical Specifications:

Model Number MC 000230
Inner dimensions - Chamber, stainless steel
 
Volume 7.8 cu ft. (222 l)
Depth 20.5" (520 mm)
Width 21.3" (540 mm)
Height 29.9" (760 mm)
Trays, stainless steel *)
 
Standard Equipment  2 pcs
Storage area (w × d) 20.5" x 19.1" (520 x 485 mm)
Racks  10 max
Distance Between guide rails 2.8" (70 mm)
Maximal weight of the load
 
Inside the oven 154 lbs. (70 kg)
Per 1 tray 66 lbs. (30 kg)
Door  1 pcs.

External Dimensions (Including Door and Handle)

Depth 31" (790 mm)
Diameter of the Air Branch Outer / Inner 2" / 1.9" (52/49 mm)
Width 29.9" (760 mm)
Height, incl. Feet (N) and Rolls (K) 43.1" (1095 mm)

Package Dimensions (Three Layers Carton)

Depth 33.9" (860 mm)
Width 33.9" (860 mm)
Depth 49.6" (1260 mm)
Weight
Brut 255 lbs. (116 kg)
Net 220 lbs. (100 kg)

Working temperature (beginning of the regulation) from 5 °C above ambient temp. to

 100°C

Temperature accuracy according to DIN 12 880 T2 at working temperature with closed air flap and door

Time variation cca (±)
0,2°C
Space deviation cca (±) 0, 3°C
Time required to reach 37 °C with closed air flap 66 min
Heat Emission at 37°C 95 w
Air Exchange speed 37°C 24 per hour 
Electricity mains 50/60 Hz
Max Power 1.8kW
Nominal Voltage 115V
Current 6A
Power stand by 690W

 

Stand Sold Separately

More Information
Technical Specifications
Model Number MC 000230
Inner dimensions - Chamber, stainless steel
 
Volume 7.8 cu ft. (222 l)
Depth 20.5" (520 mm)
Width 21.3" (540 mm)
Height 29.9" (760 mm)
Trays, stainless steel *)
 
Standard Equipment  2 pcs
Storage area (w × d) 20.5" x 19.1" (520 x 485 mm)
Racks  10 max
Distance Between guide rails 2.8" (70 mm)
Maximal weight of the load
 
Inside the oven 154 lbs. (70 kg)
Per 1 tray 66 lbs. (30 kg)
Door  1 pcs.

External Dimensions (Including Door and Handle)

Depth 31" (790 mm)
Diameter of the Air Branch Outer / Inner 2" / 1.9" (52/49 mm)
Width 29.9" (760 mm)
Height, incl. Feet (N) and Rolls (K) 43.1" (1095 mm)

Package Dimensions (Three Layers Carton)

Depth 33.9" (860 mm)
Width 33.9" (860 mm)
Depth 49.6" (1260 mm)
Weight
Brut 255 lbs. (116 kg)
Net 220 lbs. (100 kg)

Working temperature (beginning of the regulation) from 5 °C above ambient temp. to

 100°C

Temperature accuracy according to DIN 12 880 T2 at working temperature with closed air flap and door

Time variation cca (±)
0,2°C
Space deviation cca (±) 0, 3°C
Time required to reach 37 °C with closed air flap 66 min
Heat Emission at 37°C 95 w
Air Exchange speed 37°C 24 per hour 
Electricity mains 50/60 Hz
Max Power 1.8kW
Nominal Voltage 115V
Current 6A
Power stand by 690W
Incubator Type Forced Air Circulation
Controller Digital